Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Oculus Technologies of Mexico, S.A. de C.V. recalls PhaseOne Antimicrobial Solution

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
Products
  • PhaseOne Antimicrobial Solution
Company
Oculus Technologies of Mexico, S.A. de C.V.
Where it was sold

US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.

States: Arizona, California, Colorado, Florida, Georgia, Illinois, Indiana, Kentucky, Michigan, Mississippi, North Carolina, New Jersey, New York, Tennessee, Texas, Utah, Virginia

Barcodes
00860009268616
Units affected
5904 units
About this recall
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050