Shimadzu Medical Systems recalls Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.
- Products
-
- Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40
- Company
- Shimadzu Medical Systems
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, The Bahamas, China, Denmark, Haiti.
Nationwide
- Units affected
- 3487 units
- About this recall
- Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST alexa:UD150G-40 / D150GC-40; Digital Radiography System RADspeed Pro: UD150B-40 / UD150V-40 / UD150L-40 / UD150L-40E/D150BC-40 / D150VC-40 / D150LC-40 / D150LC-40E/D150BC-41 / D150VC-41 / D150LC-41; Digital Radiography System RADspeed safire: UD150B-40 / D150BC-40; Radiography System RADspeed fit: N/A (Integrated in the system); X-RAY R/F SYSTEM SONIALVISION safire17:UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION safire Plus: UD150B-40 / D150BC-4;0 X-RAY R/F SYSTEM SONIALVISION safire: UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION VERSA: UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM FLUOROspeed 300:UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM SONIALVISION G4:D150BC-40S; X-RAY R/F SYSTEM FLUOROspeed X1:D150BC-40S