PIE Medical Imaging B.V. recalls 3mensio Workstation (Vascular Fenestrated) software

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage
Products
  • 3mensio Workstation (Vascular Fenestrated) software
Where it was sold

US Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA, TX, MS, NE, NC.

States: California, Georgia, Kansas, Massachusetts, Minnesota, Mississippi, North Carolina, Nebraska, New Jersey, New York, Ohio, Texas

Units affected
501
About this recall
3mensio Workstation (Vascular Fenestrated) software