PIE Medical Imaging B.V. recalls 3mensio Workstation (Vascular Fenestrated) software
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage
- Products
-
- 3mensio Workstation (Vascular Fenestrated) software
- Company
- PIE Medical Imaging B.V.
- Where it was sold
-
US Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA, TX, MS, NE, NC.
States: California, Georgia, Kansas, Massachusetts, Minnesota, Mississippi, North Carolina, Nebraska, New Jersey, New York, Ohio, Texas
- Units affected
- 501
- About this recall
- 3mensio Workstation (Vascular Fenestrated) software