PIE Medical Imaging B.V. recalls 3mensio Workstation (Vascular Fenestrated) software
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
- Products
-
- 3mensio Workstation (Vascular Fenestrated) software
- Company
- PIE Medical Imaging B.V.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.
States: California, Michigan, Minnesota, New York, Texas
- Units affected
- 165
- About this recall
- 3mensio Workstation (Vascular Fenestrated) software