Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ICU Medical, Inc. recalls Plum Duo Infusion Pump

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Products
  • Plum Duo Infusion Pump
Where it was sold

Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.

States: Idaho, Illinois, Indiana, Kansas, Massachusetts, Maryland, New Mexico, New York, Pennsylvania, Tennessee, Texas

Units affected
13,613
About this recall
Plum Duo Infusion Pump, 40002-0401