Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Boston Scientific Corporation recalls ENDOTAK RELIANCE G/SG

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Products
  • ENDOTAK RELIANCE G/SG
Where it was sold

Worldwide Distribution. US nationwide. International distribution worldwide.

Nationwide

Barcodes
00802526372766008025263772590080252637325100802526373282008025263733740080252637346600802526373558008025263736400080252637373200802526376726008025263768180080252641197700802526411984008025264220410080252642205800802526412011008025264120280080252641203500802526412042
Units affected
247,702 US; 344,670 OUS
About this recall
ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187