Boston Scientific Corporation recalls ENDOTAK RELIANCE G/SG with 4-SITE Connector
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
- Products
-
- ENDOTAK RELIANCE G/SG with 4-SITE Connector
- Where it was sold
-
Worldwide Distribution. US nationwide. International distribution worldwide.
Nationwide
- Barcodes
0080252643137100802526431388008025264314010080252643141800802526431432008025264314490080252643146300802526431470- Units affected
- 247,702 US; 344,670 OUS
- About this recall
- ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296