Boston Scientific Corporation recalls RELIANCE 4-FRONT
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
- Products
-
- RELIANCE 4-FRONT
- Where it was sold
-
Worldwide Distribution. US nationwide. International distribution worldwide.
Nationwide
- Barcodes
008025265609340080252656095800802526560965008025265609720080252656098900802526519017008025265188500080252651909300802526519987- Units affected
- 247,702 US; 344,670 OUS
- About this recall
- RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696