Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Boston Scientific Corporation recalls RELIANCE 4-FRONT

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Products
  • RELIANCE 4-FRONT
Where it was sold

Worldwide Distribution. US nationwide. International distribution worldwide.

Nationwide

Barcodes
008025265609340080252656095800802526560965008025265609720080252656098900802526519017008025265188500080252651909300802526519987
Units affected
247,702 US; 344,670 OUS
About this recall
RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696