Cardinal Health 200, LLC recalls Presource kits and trays
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Cardinal Health is following the FDA's recommendations regarding the disposition of affected Jiangsu Shenli Medical Production Co. Ltd (Shenli) Syringes in the Merit Medical¿ supplied finished kits. Merit Medical¿ supplied finished kits are included as components in Presource kits. According to the FDA s communication, these Shenli syringes, located inside a Merit Medical¿ kit component, cannot be distributed for use as they could contain a significant change or modification in design which could significantly alter the safety or effectiveness of the finished kit product, resulting in the risk of patient harm such as inaccurate dosing and a leaking device.
- Products
-
- Presource kits and trays
- Company
- Cardinal Health 200, LLC
- Where it was sold
-
Nationwide
Nationwide
- Units affected
- 17445 units
- About this recall
- Presource kits and trays, labeled as: 1) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC3; 2) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC4; 3) Cardiovascular Procedure Kit, Catalog Number SAN29ANCCL; 4) Cardiovascular Procedure Kit, Catalog Number SAN29ANCCM; 5) Cardiovascular Procedure Kit, Catalog Number SAN33CDCAS; 6) Cardiovascular Procedure Kit, Catalog Number SAN33CDCAT; 7) Cardiovascular Procedure Kit, Catalog Number SANOCAENSA; 8) Cardiovascular Procedure Kit, Catalog Number SCV29ANCCA; 9) Left Heart Kit, Catalog Number SAN73LHKCC; 10) Left Heart Kit, Catalog Number SAN73LHKC6