Micromed S.p.A. recalls SD LTM 64 PLUS
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
- Products
-
- SD LTM 64 PLUS
- Company
- Micromed S.p.A.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of FL, PA.
States: Florida, Pennsylvania
- Barcodes
8033928120897- Units affected
- 207 units
- About this recall
- SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02