Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

DT MedTech, LLC recalls Hintermann Series Talar Implant

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Through customer complaints, the ankle replacement system was found to contain incorrect devices: Left implant packaging and labeling bears a right implant device.
Products
  • Hintermann Series Talar Implant
Company
DT MedTech, LLC
Where it was sold

Worldwide - US Nationwide distribution in the states of CA, AR, AZ, ID and the countries of PER, KOR.

States: Arkansas, Arizona, California, Idaho

Units affected
67 units
About this recall
Hintermann Series Talar Implant, Left, Size 2 REF 302112