Palodex Group Oy recalls Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-r
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).
- Products
-
- Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-r
- Company
- Palodex Group Oy
- Where it was sold
-
US Nationwide-Worldwide
Nationwide
- Units affected
- 103
- About this recall
- Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.