Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Palodex Group Oy recalls Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-r

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).
Products
  • Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-r
Company
Palodex Group Oy
Where it was sold

US Nationwide-Worldwide

Nationwide

Units affected
103
About this recall
Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.