Boston Scientific Corporation recalls WATCHMAN TruSeal Access System SGL
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
- Products
-
- WATCHMAN TruSeal Access System SGL
- Where it was sold
-
Worldwide.
Nationwide
- Barcodes
08714729965732- Units affected
- 340185 units in total
- About this recall
- WATCHMAN TruSeal Access System SGL, OUS, Material Number (UPN) M635TS70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.