Medtronic Neuromodulation recalls A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Patients implanted with a pocket adaptor (Model 64001 and/or 64002) are limited to HEAD ONLY MRI eligibility.
- Products
-
- A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584
- Company
- Medtronic Neuromodulation
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Canary Islands, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Iran, Islamic Republic Of, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic Of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Puerto Rico, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom, Uruguay, Viet Nam.
Nationwide
- Barcodes
006139945449400061399454495700613994761057006431698642140076300042098700763000564346- Units affected
- 18,010 units
- About this recall
- A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1057, 3.0.1062, 3.0.1081, and 3.0.1098), Version 4.0.x (4.0.1052). Includes the below product descriptions: 1. ADAPTOR 64001 1X4 MVMT POCKET ADAPTOR, 2. ADAPTOR 64002 2X4 MVMT POCKET ADAPTOR, 3. INS 37602 Activa SC Neurostimulator, 4. INS 37603 Activa SCx Neurostimulator, 5. INS B35200 DBS PERCEPT PC, 6. INS B35300 PERCEPT RC V1