Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Link Bio Corp recalls LINK SymphoKnee Patella Sizing Template

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.
Products
  • LINK SymphoKnee Patella Sizing Template
Company
Link Bio Corp
Where it was sold

US Nationwide distribution in the states of FL, GA, KS, OH, WI.

States: Florida, Georgia, Kansas, Ohio, Wisconsin

Barcodes
04026575310203
Units affected
8 units
About this recall
LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00