Wasatch Photonics recalls Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
- Products
-
- Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers
- Company
- Wasatch Photonics
- Where it was sold
-
U.S.
Nationwide
- Units affected
- 192
- About this recall
- Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.