Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Wasatch Photonics recalls Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
Products
  • Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers
Company
Wasatch Photonics
Where it was sold

U.S.

Nationwide

Units affected
192
About this recall
Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.