Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Lumicell, Inc. recalls Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The potential of the elastic bands included in the kit detaching or fracturing during surgical procedures. As a result, these failures may pose a risk of retained foreign bodies in the surgical cavity and potentially leading to surgical site infections.
Products
  • Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack
Company
Lumicell, Inc.
Where it was sold

US Nationwide distribution in the states of CA, FL, MA, NC, TX.

States: California, Florida, Massachusetts, North Carolina, Texas

Units affected
200 units (20 packs containing 10 units per pack)
About this recall
Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Model/Catalog Number: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Software Version: Not Applicable Product Description: Sterile Cover Component: Lumicell Direct Visualization System (DVS) / PMA # P230014