FDAMedical devicesRecalled Status: Ongoing

Pfm Medical Inc. recalls Multi-Snare Set: 5 mm x 125 cm

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
Products
  • Multi-Snare Set: 5 mm x 125 cm
Company
Pfm Medical Inc.
Where it was sold

US Nationwide distribution in the states of NY, MD.

States: Maryland, New York

Units affected
32
About this recall
Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310