Pfm Medical Inc. recalls Multi-Snare Set: 5 mm x 125 cm
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
- Products
-
- Multi-Snare Set: 5 mm x 125 cm
- Company
- Pfm Medical Inc.
- Where it was sold
-
US Nationwide distribution in the states of NY, MD.
States: Maryland, New York
- Units affected
- 32
- About this recall
- Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310