Boston Scientific Corporation recalls Boston Scientific NephroMax Kit
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.
- Products
-
- Boston Scientific NephroMax Kit
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
0871472907758908714729834540- Units affected
- 458 units
- About this recall
- Boston Scientific NephroMax Kit, Material Numbers (UPN): 1. M0062101180; 2. M0062101600. (Urology)