Medacta Usa Inc recalls Short Monobloc Stem Broaches
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
- Products
-
- Short Monobloc Stem Broaches
- Company
- Medacta Usa Inc
- Where it was sold
-
US Nationwide distribution in the states of CO, AR, IN, PA, UT, FL.
States: Arkansas, Colorado, Florida, Indiana, Pennsylvania, Utah
- Units affected
- 50
- About this recall
- Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System