Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Zap Surgical Systems recalls ZAP-X Radiosurgery System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
Products
  • ZAP-X Radiosurgery System
Company
Zap Surgical Systems
Where it was sold

Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.

States: Colorado, Florida, New Jersey, New York

Barcodes
00860183001504
Units affected
17
About this recall
ZAP-X Radiosurgery System, Model: 300150