Zap Surgical Systems recalls ZAP-X Radiosurgery System
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
- Products
-
- ZAP-X Radiosurgery System
- Company
- Zap Surgical Systems
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.
States: Colorado, Florida, New Jersey, New York
- Barcodes
00860183001504- Units affected
- 17
- About this recall
- ZAP-X Radiosurgery System, Model: 300150