Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
- Products
-
- MiniMed 780G Insulin Pump
- Company
- Medtronic MiniMed, Inc.
- Where it was sold
-
International distribution to the countries of Spain, Great Britain, Netherlands, Italy.
- Units affected
- 778
- About this recall
- MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896