Wallac Oy recalls GSP Neonatal GALT Kit. Product Codes: 3303-001U
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Reduction in the on-board stability of the GSP GALT assay. This may increase the likelihood of run-QC failures and consequently delay the reporting of patient results.
- Products
-
- GSP Neonatal GALT Kit. Product Codes: 3303-001U
- Company
- Wallac Oy
- Where it was sold
-
Worldwide - US Nationwide distribution in the states and territories of AL, AR, AZ, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NJ, OR, PR, SC, TN, TX, VA, WV and the countries of Austria, Belgium, Egypt, France, Germany, Hong Kong, Iraq, Italy, Netherlands, Oman, Philippines, Saudi Arabia, Singapore, Spain, Thailand, United Arab Emirates, Vietnam.
States: Alabama, Arkansas, Arizona, Colorado, Florida, Georgia, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Maryland, Minnesota, Missouri, New Jersey, Oregon, Puerto Rico, South Carolina, Tennessee, Texas, Virginia, West Virginia
- Barcodes
0643814732170406438147311347- Units affected
- 3,274 units
- About this recall
- GSP Neonatal GALT Kit. Product Codes: 3303-001U, 3303-0010. in vitro diagnostic. Used in FLUORESCENT PROC. (QUAL), GALACTOSE-1-PHOSPHATE URIDYL TRANSFERASE.