Olympus Corporation of the Americas recalls Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
- Products
-
- Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America.
Nationwide
- Barcodes
00821925003538- Units affected
- 3,663 units
- About this recall
- Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.