Tosoh Bioscience Inc recalls ST AIA-Pack Testosterone
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.
- Products
-
- ST AIA-Pack Testosterone
- Company
- Tosoh Bioscience Inc
- Where it was sold
-
Global Distribution. US Nationwide.
Nationwide
- Barcodes
14560189210032- Units affected
- 310 units
- About this recall
- ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204