Boston Scientific Corporation recalls FARADRIVE STEERABLE SHEATH CLEAR OUS
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
- Products
-
- FARADRIVE STEERABLE SHEATH CLEAR OUS
- Where it was sold
-
Global Distribution. US Nationwide.
Nationwide
- Barcodes
00191506042516- Units affected
- 193,000 units
- About this recall
- FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01