Medtronic Perfusion Systems recalls EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
- Products
-
- EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR
- Company
- Medtronic Perfusion Systems
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
2076300013563200763000135638- Units affected
- 21,290 units
- About this recall
- EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422