Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medtronic Perfusion Systems recalls EOPA 3D Arterial Cannula DIL TIP 22FR

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Products
  • EOPA 3D Arterial Cannula DIL TIP 22FR
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
2076300013570000763000135706
Units affected
3,621 units
About this recall
EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222