Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Bioporto Diagnostics A/S recalls ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.
Products
  • ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.
Company
Bioporto Diagnostics A/S
Where it was sold

Worldwide - The affected product distribution included the United States and the countries of European Union, Chile, and Hong Kong. No affected product was distributed to Canada; however, Health Canada was notified because the product is registered in Canada.

Nationwide

Barcodes
5711435000156
Units affected
400 units
About this recall
ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.