Philips Medical Systems Nederland B.v. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.
Products
  • Philips Azurion 7M20 systems with FlexArm ceiling-mounted system
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
00884838085268
Units affected
927 systems
About this recall
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224