Philips Medical Systems Nederland B.v. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.
- Products
-
- Philips Azurion 7M20 systems with FlexArm ceiling-mounted system
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00884838085268- Units affected
- 927 systems
- About this recall
- Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224