Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Tobii Dynavox Llc recalls TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Products
  • TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Where it was sold

US Nationwide distribution.

Nationwide

About this recall
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798