Tobii Dynavox Llc recalls TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
- Products
-
- TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
- Company
- Tobii Dynavox Llc
- Where it was sold
-
US Nationwide distribution.
Nationwide
- About this recall
- TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799