Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Paragonix Technologies Inc. recalls Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Products
  • Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
Company
Paragonix Technologies Inc.
Where it was sold

US Nationwide distribution in the states and territories of AL, AZ, CA, TN, FL, HI, MN, NJ, TX, UT, CO, CT, DC, MD, GA, KY, LA, MA, MI, MO, MS, NC, ND, NE, NM, NV, NY, PA, PR, SC, UT, VA, WI, WV, IN, OH, KY, IL, MD, MN, OR.

States: Alabama, Arizona, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, Nebraska, New Jersey, New Mexico, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia

Barcodes
00850005470342
Units affected
1445 systems
About this recall
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001