GE Medical Systems Information Technologies Inc recalls PORTRAIT COMPUTE AND CORE SERVICES V1.2
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
- Products
-
- PORTRAIT COMPUTE AND CORE SERVICES V1.2
- Where it was sold
-
US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.
States: Michigan, New York
- Barcodes
00198953056752- Units affected
- 4 units
- About this recall
- PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592