GE Medical Systems Information Technologies Inc recalls
GE Medical Systems Information Technologies Inc is named on 9 recalls in Recall Relay's records since November 2023, from FDA (17 recalled products: FDA lists each product as its own recall); the most recent was July 30, 2026.
All 17 recalled products, newest first.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.1Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls FRU (Field replaceable unit)Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls PORTRAIT COMPUTE AND CORE SERVICES V1.2Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.0.5Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE Healthcare Portrait Core Services SW V 1.2.0A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or access to Protected Health Information (PHI).
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE Healthcare MAC VU360Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODELGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service onlyGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVERGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE HealthCare CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICAGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE HealthCare ApexPro v6.0 eDelivery SoftwareGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICAGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls MUSE 5 SystemsThe MUSE Administrator Account active directory or local credential could potentially be used from your local network to log into a GE HealthCare MUSE 5 system and potentially manipulate recorded patient data.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AlCapacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.
-
Class IIofficial
GE Medical Systems Information Technologies Inc recalls GE HealthCare CRITIKON ONE-CUF blood pressure cuffsUse of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.