Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Seer Medical Pty Ltd recalls Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the current product due to design changes in the system. A new CAPA was opened to investigate the radiated emissions compliance issues. - Seer Home Systems distributed and retained in the field were not compliant with Electromagnetic Compatibility (EMC) testing according to IEC 60601-1-2 - Seer Home Systems, including accessories, were manufactured and distributed without application of proper design change and manufacturing controls - Seer Home Systems were distributed with labeling deficiencies
Products
  • Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG
Company
Seer Medical Pty Ltd
Where it was sold

Worldwide - US Nationwide distribution in the states of MN, NY, PA and the countries of Australia.

States: Minnesota, New York, Pennsylvania

Barcodes
09355751001037093557510010200935575100113609355751001112093557510011290935575100109909355751001105
Units affected
3,948 units
About this recall
Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory components (Disposable EEG Wires, WaterTabs, WaterTabs Application Kit, and Take-home Kit).