Btt Health Gmbh recalls Marodyne-LiV
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Some power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).
- Products
-
- Marodyne-LiV
- Company
- Btt Health Gmbh
- Where it was sold
-
US Nationwide distribution in the states of Florida.
States: Florida
- Units affected
- 62 units
- About this recall
- Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment