Datex-Ohmeda, Inc. recalls Aisys
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
- Products
-
- Aisys
- Company
- Datex-Ohmeda, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00840682102292- Units affected
- 475 units
- About this recall
- Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000