Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Products
  • Aisys CS2 with Et Control
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Units affected
7 units
About this recall
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000