Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Howmedica Osteonics Corp. recalls DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M0

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Products
  • DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M0
Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

Nationwide

Barcodes
0761332713819107613327138214076133271382070761332713815307613327138283076133271381770761332713826907613327138238
Units affected
12,156 US
About this recall
DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-000 SHOULDER - GLENOID #11 5361-6111 SHOULDER - GLENOID #5 5361-6105 SHOULDER - GLENOID #7 5361-6107 SHOULDER - GLENOID #9 5361-6109 SHOULDER - KEELED GLENOID #11 5361-6211 SHOULDER - KEELED GLENOID #5 5361-6205 SHOULDER - KEELED GLENOID #7 5361-6207 SHOULDER - KEELED GLENOID #9 5361-6209 Orthopaedic implant components manufactured from UHMWPE (polyethylene)