American Contract Systems, Inc. recalls Essentia MAJ Laparotomy Std Pack
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
- Products
-
- Essentia MAJ Laparotomy Std Pack
- Where it was sold
-
US Nationwide distribution in the states of MO, MN, MA, OH, NE.
States: Massachusetts, Minnesota, Missouri, Nebraska, Ohio
- Barcodes
00191072218438- Units affected
- 68 units
- About this recall
- Essentia MAJ Laparotomy Std Pack, ESLT53B; Medical convenience kit