Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Healgen Scientific Llc recalls Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by way of uneven color fading on the total protein test pad. This could lead to a delay in the identification of potential kidney diseases, their evaluation, and/or other conditions associated with elevated protein levels in urine and subsequently a delay in patient treatment.
Products
  • Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each
Company
Healgen Scientific Llc
Where it was sold

U.S. Nationwide distribution in the states of CA, GA and TX.

States: California, Georgia, Texas

Barcodes
196852593453008164900243470996540110430081649002432370748698753470748698705300816490024330
Units affected
1,636,994 Tests
About this recall
Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing one assay item by visually or instrumentally reading the color change of the pad and comparing with the corresponding blocks on a color chart. The Healgen Series Reagent Strips for Urinalysis provides tests for Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Ascorbic Acid and Leukocytes in Urine.