Aizu Olympus Co., Ltd. recalls Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
Class IIofficial
Olympus Corporation of the Americas recalls Insufflation unitOlympus has become aware of an increased trend of both repairs and customer complaints reporting "UHI-4 stopping air supply due to the front panel LED turning off".
-
Class IIofficial
Townsend Design recalls CUSTOM SPRYSTEPDue to potential air pockets between layers of material, their is the potential for complete struct breakage of knee-ankle-foot orthosis.
-
Class IIofficial
Townsend Design recalls SPRYSTEP VECTORDue to potential air pockets between layers of material, their is the potential for complete struct breakage of knee-ankle-foot orthosis.
-
Class IIofficial
Townsend Design recalls SPRYSTEP VECTOR KAFODue to potential air pockets between layers of material, their is the potential for complete struct breakage of knee-ankle-foot orthosis.
-
Class IIofficial
Musculoskeletal Transplant Foundation, Inc. recalls DBX Demineralized Bone Matrix Putty 10cc-Indicated for treatment of surgically created osseous defects or osseUnits are labeled as 10cc; however the units have a fill volume of 1cc units. This may result in prolongation of surgery.
-
Class IIofficial
Philips North America Llc recalls CombiDiagnost R90Potential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls DigitalDiagnost 4 Chest/EmergencyPotential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls DigitalDiagnost 4 Flex/ValuePotential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls DigitalDiagnost 4 High PerformancePotential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls DigitalDiagnost C90 Flex/Value/Chest/ERPotential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls DigitalDiagnost C90 High PerformancePotential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls DigitalDiagnost C90 Release 3Potential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls DigitalDiagnost C90 Release 3.1Potential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls EasyDiagnost Eleva DRFPotential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Philips North America Llc recalls ProxiDiagnost N90Potential for units suspended on the ceiling to fail and fall.
-
Class IIofficial
Atrium Medical Corporation recalls Atrium Advanta V12 Covered Stent SystemBalloon-to-catheter weld failure that may cause loss of balloon pressure during stent deployment.
-
Class IIofficial
ROi CPS LLC recalls Regard CV PKSpecific Regard kits contain the Gentige Chest Drain Products being recalled by Gentige. The Instructions for Use (IFU) for the chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains.
-
Class IIofficial
Magellan Diagnostics, Inc. recalls LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measureDuring an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
-
Class IIofficial
NeuroSync, Inc. recalls NeuroSync EYE-SYNC containing Pico Neo 2 Headsets. Each unit consists of a Pico Neo 2 headset with embedded seEye syncing issues causing the device to be inoperable. Unable to be repaired due to discontinuation of support for the Pico 2 platform, which was no longer in production or serviceable as of August 31, 2023.
-
Class IIofficial
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235. PTS-11 versionsThe patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
-
Class IIofficial
Atrium Medical Corporation recalls Ocean Water Seal Chest Drains (Ocean): Products REF Number Product Name (1)2002-000 DRAINInstructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains
-
Class IIofficial
Atrium Medical Corporation recalls Oasis Dry Suction Water Seal Chest Drains (Oasis): (1)3600-100 DRAINInstructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains
-
Class IIofficial
Atrium Medical Corporation recalls Express Dry Seal Chest Drains (Express): 4000-100N DRAINInstructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains
-
Class Iofficial
CareFusion 303, Inc. recalls BD Alaris PCU REF 8015Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
-
Class Iofficial
CareFusion 303, Inc. recalls BD Alaris Syringe ModuleAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
-
Class Iofficial
CareFusion 303, Inc. recalls Alaris PCA Module 8120Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
-
Class IIofficial
Olympus Corporation of the Americas recalls EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera headsCertain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.
-
Class IIofficial
Hologic, Inc. recalls Aptima CMV Quant AssayUse of assay along with plasma specimen may result in error code which would invalidate sample.
-
Class IIofficial
Breg Inc recalls BREG REF 100519-000 WALKER FOLDING WALKER WITH WHEELS Utilized as a walking aid.Due to walkers not meeting product specifications. Walkers were manufactured with misaligned wheels.
-
Class IIofficial
Stryker Corporation recalls StrykerExpired product distributed
-
Class IIofficial
Ortho-Clinical Diagnostics, INc. recalls VITROS Chemistry Products Lipase SlidesProducts 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
-
Class IIofficial
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forcepsAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.