Fisher & Paykel Healthcare, Ltd. recalls
Fisher & Paykel Healthcare, Ltd. is named on 3 recalls in Recall Relay's records since November 2023, from FDA; the most recent was September 18, 2025.
All 3, newest first.
-
Class IIofficial
Fisher & Paykel Healthcare, Ltd. recalls Airvo 2When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.
-
Class IIofficial
Fisher & Paykel Healthcare, Ltd. recalls PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended toDue to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.
-
Class IIofficial
Fisher & Paykel Healthcare, Ltd. recalls Airvo 2 HumidifierHumidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Fisher & Paykel Healthcare, Ltd. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.