Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Fisher & Paykel Healthcare, Ltd. recalls Airvo 2 Humidifier

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
Products
  • Airvo 2 Humidifier
Where it was sold

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.

Nationwide

Units affected
7,147
About this recall
Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US