Fisher & Paykel Healthcare, Ltd. recalls PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.
Products
  • PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to
Where it was sold

U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided

States: Arizona, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Iowa, Illinois, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Montana, North Dakota, New Hampshire, New Jersey, Nevada, New York, Ohio, Oregon, Pennsylvania, South Dakota, Texas, Utah, Virginia, Washington

Barcodes
09420012466662
Units affected
944 units
About this recall
PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.