FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
- Products
-
- FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull
- Where it was sold
-
US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.
States: Alaska, California, Florida, Illinois, Nebraska, New Hampshire, Ohio, Tennessee, Wisconsin
- Barcodes
04540217052226- Units affected
- 18 units
- About this recall
- FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33