Shimadzu Corp. Analytical & Measuring Instruments Div. recalls
Shimadzu Corp. Analytical & Measuring Instruments Div. is named on 2 recalls in Recall Relay's records since November 2023, from FDA; the most recent was October 9, 2024.
All 2, newest first.
-
Class IIofficial
Shimadzu Corp. Analytical & Measuring Instruments Div. recalls AutosamplerThere have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.
-
Class IIofficial
Shimadzu Corp. Analytical & Measuring Instruments Div. recalls AutosamplerThere have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.