Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
- Products
-
- Autosampler
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
States: Delaware, Pennsylvania
- Units affected
- 48
- About this recall
- Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL