Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.
- Products
-
- Autosampler
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.
States: Arkansas, California, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Kentucky, Massachusetts, Michigan, Minnesota, Missouri, Mississippi, New Jersey, Nevada, New York, Ohio, Oklahoma, South Dakota, Texas, Utah, Virginia, Washington, Wisconsin
- Units affected
- 179
- About this recall
- Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL